Model : ST-HT60P; ST-HT100P; ST-HT130P; ST-HT220P
Capacity (L) : 60 / 108 / 135 / 192
Product Form : Automatic door vertical
Select Sterilizer : 58%-60% hydrogen peroxide (H2O2)
Working Temperature (℃) : 45-60
Sterilization Time (min) : 35-55
The hydrogen peroxide low temperature plasma sterilizer is used in medical facilities, laboratories and the pharmaceutical industry, among others, for a wide range of effective uses.
You can later choose whether you want a quote, or if you place a firm order.
Product Description
Low temperature hydrogen peroxide plasma sterilizers are used for sterilizing heat-sensitive medical devices, dental equipment, laboratory equipment, pharmaceuticals, veterinary instruments, and beauty industry tools. They provide safe and effective sterilization while preserving delicate materials.
Hydrogen Peroxide Gas Plasma Sterilization Technical Features:
- Low temperature sterilization, preserving delicate items without damage.
- Green and pollution-free process using hydrogen peroxide solution, no toxic substances or pollution.
- Stable sterilization with balanced temperature and stable hydrogen peroxide concentration.
- Efficient design with rectangular sterilization chamber for increased loading capacity.
- Safe and stable operation with electric lift door and safety interlock device.
- Simple operation with intelligent control processor and one-key start.
- Configured with high-quality components for reliable performance.
- Time and cost-saving with short sterilization cycle and low operating costs.
- With sterile filter for air inlet to prevent secondary pollution.
- U disk storage function for storing and recording sterilization data.
Low Temp H₂O₂ Plasma Sterilizer Specifications:
Low Temp H₂O₂ Plasma Sterilizer Features:
- High-quality 5052 aluminum material for sterilization room, with excellent thermal conductivity.
- ABS plastic disposable outer box material.
- Aluminum alloy electrode mesh and sealing door material.
- Upper and lower layers of aluminum alloy punching plates for loading sterilized items.
- Electric lifting door with infrared sensor switch and door anti-pinch function.
- Preheating heating time of ≤30 minutes.
- Full automatic control system with LCD touch screen display and PLC circuit design.
- Micro-printer for printing and saving data during sterilization.
- Siemens smart series PLC control system.
- 7-inch color touch screen display with various status information.
- Storage function for recording sterilization data.
- Built-in pressure and temperature sensors for precise control.
- Two program options for circulation and rapid circulation.
- Network connection port available.
- Vacuum system with fast vacuum rate and long service life.
- Oil mist filter and air filter for exhaust filtration.
- Transistor control power supply for stable power output.
- Cartridge filling method for hydrogen peroxide.
- Explosion-proof flame retardant heating system.
- Over-voltage and over-temperature protection functions.